Chapter 9: Calibration / adjustment
9-16 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
9.1.18 Part 10: Calibrating the 800 flow rate
9.1.19 Part 11: Calibrating the dialysate temperature
Note
Accuracy of the measuring instrument to be connected externally:
±0.2 °C.
CAL. FLOW 800 ml/min
flow (800) = XXX
DATA STORED
Conf
key
Audio paused
key
Esc
key
Change the digits in display
(field ) by pressing the +/– keys,
until the actual value agrees
with the specified value (500).
After
approx. 3 s
CAL. FLOW 800 ml/min
xxx
DAC digits




CALIB. TEMPERATURE
Temp. ADJUSTMENT
Conf
key
see Part 11.1
back to menu ?
Check TEMPERATURE see Part 11.2
Conf
key
Conf
key
Conf
key
OCM TEMP.-Check see Part 11.3