Chapter 11: Functional description
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 11-17
11.1.1.16 UF function test
Test description – CPU1 activates the UF pump with a defined rate.
– CPU2 checks the UF pump.
– CPU2 disables the control line of the UF pump and checks it for
standstill.
– Check the UF counter.
+P_DIAL
X633R/
C28
UF_P1
UF_P_EN
Rückmeldung/Acknowledgement
Testgenerierung/Generation of Test