Chapter 9: Calibration / adjustment
9-22 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
9.1.26 Part 12.3: Calibrating the concentrate pump stroke
Note
Measuring cylinder accuracy: ±0.5 %.
cal. CONC.-PUMP-VOL
Conf
key
Audio paused
key
CONP-Vol = 0.828 ml
Enter the determined
volume of one pump stroke
by pressing the +/– keys.
DATA STORED
Esc
key
After
approx. 3 s