Chapter 9: Calibration / adjustment
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 9-15
Turn potentiometer P1 in clockwise direction until the device
switches from intrinsic pulse mode ("Eigentakt") to regular balancing
chamber switching mode (wait for approx. 10 seconds after each
revolution!).
–Display: flow (300) = XXX
If necessary, correct the flow with the + / – keys until the actual flow
is approx. 300.
Turn potentiometer P1 in anti-clockwise direction (wait for at least
10 seconds after each revolution!) until the device switches to
intrinsic pulse mode ("Eigentakt").
–Display: flow (300) = 147
Turn potentiometer P1 in clockwise direction (wait for at least
10 seconds after each half revolution!) until the actual flow is again
approx. 300.
Turn potentiometer P1 in clockwise direction for another
2 revolutions.
9.1.17 Part 9: Calibrating the 500 flow rate
Note
After having adjusted the current increasing pulse, check and, if
necessary, readjust the 300 / 500 / 800 flow rate settings.
CAL. FLOW 500 ml/min
flow (500) = XXX
DATA STORED
Conf
key
Audio paused
key
Esc
key
Change the digits in display
(field ) by pressing the +/– keys,
until the actual value agrees
with the specified value (500).
After
approx. 3 s
CAL. FLOW 500 ml/min
xxx
DAC digits

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals