Chapter 7: Error messages
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 7-19
7.2.2 PSW (pressure switch) monitoring during free rinsing
(only with devices with CDS)
Error message Description
Rinse Failure F05 Rinse-free program with following Dis or HDIS or mandatory rinse as individual
program in Dis I V.
It was impossible to open the pressure switch for PSW_104 (S124)
(bicarbonate).
Pressure on distribution piping > 500 mbar (according to specification, the
permissible pressure is max. 500 mbar). Pressure peaks on distribution
piping: Frequently occurs in distribution pipings with user points if e.g.,
several patients are disconnected simultaneously and disinfection is started.
Switching point of pressure switch too low:
Desired= 700 mbar ±20 mbar
Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Concentrate: X633L/A20
Rinse Failure F06 Rinse-free program with following Dis or HDIS or mandatory rinse as individual
program in Dis I V.
It was impossible to open the pressure switch for PSW_102 (S123)
(concentrate).
Pressure on distribution piping > 500 mbar (according to specification, the
permissible pressure is max. 500 mbar). Pressure peaks on distribution
piping: Frequently occurs in distribution pipings with user points if e.g.,
several patients are disconnected simultaneously and disinfection is started.
Switching point of pressure switch too low:
Desired= 700 mbar ±20 mbar
Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Concentrate: X633L/A20
Rinse Failure F07 Rinse-free program, Dis, HDIS, or mandatory rinse in Dis I to V.
Pressure drop during the monitoring phase on PSW_104 (S124) (bicarbonate)
or pressure build-up impossible.
Check switching point of pressure switch
Check loading pressure
(possibly splinter or contamination in orifice 151, remove and purge the
tubing from both ends)
Check negative pressure and orifice (89)
(for this purpose, remove and purge the tubing from both ends)
Check CDS valve (104)
Check tightness of CDS path
Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Bicarbonate: X633L/A19
Cartridge filter (F210) upstream of degassing pump clogged or wrong filter
(filter for disinfectant container) installed. Filters can be distinguished by
different adapters.

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals