Chapter 7: Error messages
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 7-9
7.1.10 Blood leak test
7.1.11 Temperature test
Error message Description
F01 Blood Leak Blood leak channel and dimness not in alarm-free condition during the T1 test.
Dimness channel contaminated (calcium precipitate, etc.)
Acknowledgement (BLL, X637/A18) X633L/B10 voltage value within the
alarm tolerances ( 3V / 8V)
Acknowledgement (BLL_DIM, X637A/21) X633L/B11 voltage value within
the alarm tolerances ( 1.5 V / 8V)
DAC_BLL or DAC_DIM not within the tolerances (check calibration)
F02 Blood Leak The blood leak alarm / dimness alarm is not recognized during test detuning.
Acknowledgement (BLL, X637/A18) X633L/B10 voltage value not within
the alarm tolerances ( 3V)
Acknowledgement (BLL_DIM, X637A/21) X633L/B11 voltage value not
within the alarm tolerances ( 1.5 V)
Test detuning (BLL_DET, X632/A25) X633L/B27 not 5 V
Calibration of DAC_BLL or DAC_DIM is too high
Detuning (DAC_DIM, X634R/A11) X633L/C3 impossible
Dimness calibration is set to potentiometer calibration
(BR6 from pos. 1/2 to 2/3)
F03 Blood Leak After test detuning, the blood leak channel and dimness fail to enter the
alarm-free state.
Dimness channel contaminated (calcium precipitate, etc.)
Acknowledgement (BLL, X637/A18) X633L/B10 voltage value within the
alarm tolerances (3 V / 8V)
Test detuning (BLL_DET, X632/A25) X633L/B27 not 0 V
Acknowledgement (BLL_DIM, X637A/21) X633L/B11 voltage value within
the alarm tolerances ( 1.5 V / 8V)
DAC_BLL or DAC_DIM not within the tolerances (check calibration)
F95 Blood Leak System error
Error message Description
F01 Temperature The temperature measuring range is not set to hemodialysis.
Control line (HOTRINSE, X634R/C24) X639/A20 not 0 V
Acknowledgement (HOTRINSE, X634R/C24) X632/A26 not 0 V
F02 Temperature The actual temperature is less than 35.0 °C (test running time > 15 minutes).
Calibrate the temperature
The heater rod has failed
Acknowledgement (T_DIAL1, X633L/B16) X632/A24, voltage got stuck.
F03 Temperature The actual temperature is higher than 39.0 °C (test running time 15> 15
minutes).
Calibrate the temperature
The regulating sensor (NTC-2) is defective
Acknowledgement (T_DIAL1, X633L/B16) X632/A24, voltage got stuck.

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals