Chapter 7: Error messages
7-10 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
7.1.12 Negative pressure test (negative pressure holding test)
F04 Temperature The temperature failed to stabilize within 15 minutes.
Acknowledgement (T_DIAL1, X633L/B16) X632/A24 is steadily changing
(change > 0.3 °C/15 sec).
F05 Temperature Detuning in positive direction not higher than 3 °C (10 seconds).
Acknowledgement (T_DIAL1, X633L/B16) X632/A24 change in voltage
insufficient.
Detuning (T_DETADJ, X632/A23) X633R/C21 insufficient
F06 Temperature The monitor sensor indicates a constant value.
–NTC-3 defective
F07 Temperature The test release is missing (max. test running time is 10 minutes).
Run-time problem (software).
F08 Temperature CPU1 failed to transmit a bibag
®
status message within 3 seconds.
Run-time problem (software).
F09 Temperature bibag
®
NTC_BIB detuning not higher than 1 °C.
Acknowledgement (NTC_BIB, X633R/C15) ADW on P.C.B. LP 633-5,
change in voltage insufficient
Detuning (BIBAG_TE, X634R/A13) X633R/A20 insufficient
F10 Temperature bibag
®
temperature display outside of measuring range (15 to 45 °C)
Acknowledgement (NTC_BIB, X633R/C15) ADW on P.C.B. LP 633-5
F95 Temperature System error
Error message Description
F01 neg. Pressure During the start phase a negative pressure of more than 450 mmHg has
developed (max. test running time 120 seconds).
The hydraulic system is contaminated,
the air separation pump started running
F02 neg. Pressure Setting the dialysate pressure to the test pressure (–300 mmHg to –450 mmHg)
was not possible (max. test running time 120 seconds).
Upon repetition of measurement, the range was extended from
–260 mmHg to –490 mmHg.
Leakage in the hydraulic system
The UF pump is defective
F03 neg. Pressure The working point (116 digits) of the differential amplifier cannot be set correctly
(max. test running time 120 seconds).
Pressure variations are too large
The D-A converter (IC11) on P.C.B. LP 632 is defective
The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective
Acknowledgement (P_DIAL, X633L/B6) X632/A29 is defective
CI signal is missing (P.C.B. LP 632 X632/B22).
Error message Description

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals