Chapter 7: Error messages
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 7-35
7.4.4 BPM module error messages
Error message Description
BPM: Measurement value
outside the alarm limits
Blood pressure cuff not correctly connected. Leakage in pressure tubing. Alarm
limits not adjusted to the patient.
BPM: Measurement not
successful
It was not possible to determine a measurement value.
BPM: Option not available BPM option not installed.
Due to the error FXX BPM error the BPM option is no longer available during
the current treatment.
E02 BPM error
(Remove the cuff from the
patient!)
Module does not respond anymore
E03 BPM error
(Remove the cuff from the
patient!)
Module reset
E06 BPM error
(Remove the cuff from the
patient)
BPM Rcv buffer vent
BPM: Cuff cannot be inflated. CPU-P1 pressure < 25 mmHg 15 seconds after the start of a measurement
BPM: Inflation time limit
exceeded
Inflation time > 70 seconds
BPM: Cuff pressure lower
than preselected pressure
CPU-P1 pressure < 40 % of the desired pressure
BPM: Measuring time limit
exceeded
Measurement time > 175 seconds
BPM: Maximum cuff pressure
exceeded
CPU-P1 pressure > 320 mmHg
BPM: Inflation pressure
reached too quickly.
Inflation pressure reached too quickly.
Exx BPM error
(Remove the cuff from the
patient!)
Sample array overflow
E25 BPM error
(Remove the cuff from the
patient!)
PulsCount array overflow
E26 BPM error
(Remove the cuff from the
patient!)
StepCount array overflow
E70 BPM error
(Remove the cuff from the
patient!)
Auto-zero error.
A failure is detected if the CPU-P1 pressure is > ±9 mmHg or the pressure
change is > ±1 mmHg 5 seconds after starting a measurement.