Chapter 7: Error messages
7-36 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
E72 BPM error
(Remove the cuff from the
patient!)
Pressure difference between CPU-P1 and CPU-P2 on 1st stage > 20 mmHg
E73 BPM error
(Remove the cuff from the
patient!)
Vcc-CPU-P2 error
E77 BPM error
(Remove the cuff from the
patient!)
RetryCount array overflow
E78 BPM error
(Remove the cuff from the
patient!)
CPU-P2 time stamp error
E79 BPM error
(Remove the cuff from the
patient!)
Unknown error occurred
E80 BPM error
(Remove the cuff from the
patient!)
EEPROM data invalid
E81 BPM error
(Remove the cuff from the
patient!)
I²C bus hardware or checksum error
E82 BPM error
(Remove the cuff from the
patient!)
CRC error of the CPU-P2 program memory
F83 BPM error
(Remove the cuff from the
patient!)
CPU-P1 pressure zero point drift after T1 test
E84 BPM error
(Remove the cuff from the
patient!)
CPU-P2 pressure zero point drift after T1 test
E85 BPM error
(Remove the cuff from the
patient!)
AD values during calibration not acceptable (offset / gradient)
E86 BPM error
(Remove the cuff from the
patient!)
CPU-P2 pressure > 300 mmHg for more than 15 seconds
E87 BPM error
(Remove the cuff from the
patient!)
Interval pause < 30 seconds
E90 BPM error
(Remove the cuff from the
patient!)
CPU-P2 pressure > 330 mmHg for more than 300 ms.
Error message Description