Chapter 10: Servicing / repair
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 10-1
10 Servicing / repair
10.1 Precautions for working on the hemodialysis device
Installing and removing
components
Unless described otherwise, installation of components is in reverse
component removal order.
To be observed after
working on the
hemodialysis device
A disinfection, a T1 test and a check of the electrical safety must be
performed after working on the hemodialysis device.
To be observed after
stopping a disinfection
program
After a disinfection program has been stopped or if the device is to be
preserved, the hemodialysis device must be disconnected from the
water supply after a maximum of 3 days.
Warning
When the device is returned to use, check that the pressure of the water
supply meets the prescribed minimum pressure.