Chapter 5: Setup / service program
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 5-23
5.1.3.17 Part 16: Setting the rinse volume
5.1.3.18 Part 17: T1 test autostart
5.1.3.19 Part 18: Setting the venous alarm window
SET RINSE-VOLUME
RINSE-VOL: 1000 ml
DATA STORED
Conf
key
Audio paused
key
Set the desired rinse volume
(0 to 5000 ml
in 100 ml increments)
by pressing the +/– keys.
Esc
key
After
approx. 3 s
+/–
+/–
T1-TEST AUTOSTART
T1-T. Autostart: OFF
Conf
key
Esc
key
DATA STORED
T1-T. Autostart: ON
Esc
key
DATA STORED
After
approx. 3 s
After
approx. 3 s
Audio
paused
key
Audio
paused
key



VENOUS LIMITS
ASYM. LIMITS
Conf
key
VEN. WINDOW POSITION
see Part 18.1
back to menu ?
Conf
key
Conf
key
Conf
key
see Part 18.2