Chapter 9: Calibration / adjustment
9-26 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
9.1.30 Part 12.7: Checking the concentrate and bicarbonate volumes
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
check CONC/BIC VOL.
START MAND. FILLING?
Conf
key
Press ALARMTONE MUTE
ACKNOWLEDGED
Audio paused
key
Audio paused
key
MAND. FILLING active
After 9 s
mem puls left = 50
Esc
key
Esc
key
Esc
key
Esc
key
pulse-amount = 50
Audio paused
key
Esc
key
The pump runs for
50 strokes.
The display counts down.
Then determine and record
the volume removed
by the pump and compare it
to the calculated value.
Notes:
This test step permits verification of the concentrate
or bicarbonate pump volumes in accordance with
the parameters entered for the mixing system
(mixing ratio, BC volume, conc. and bic. pump volume).
The pump whose suction tube
is pulled off is activated.
50 strokes are factory-set. This setting can be
changed by pressing the +/– keys
(depending on the graduated cylinder used).
However, when returning to “CAL. MIXING-SYSTEM”,
the display will indicate the factory setting again.
Esc
key
Time left displayed
in display field
After
approx. 3 s

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals