Chapter 9: Calibration / adjustment
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 9-23
9.1.27 Part 12.4: Measuring the volume of the concentrate pump in liters
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
CONC. PUMP VOL det.
START MAND. FILLING?
Conf
key
Press ALARMTONE MUTE
ACKNOWLEDGED
Audio paused
key
Audio paused
key
MAND. FILLING active
Time left displayed
in display field
After 9 s
conc puls left = 100
Esc
key
Esc
key
Esc
key
Esc
key
conc pulses = 100
Audio paused
key
Esc
key
The concentrate pump
runs for 100 strokes.
The display counts down.
Then determine and record
the volume removed
by the pump.
Notes:
100 strokes are factory-set. This setting can be
changed by pressing the +/– keys
(depending on the graduated cylinder used).
However, when returning to “CAL. MIXING-SYSTEM”,
the display will indicate the factory setting again.
Check the volume and, if necessary,
repeat the procedure.
Esc
key
After
approx. 3 s