Chapter 10: Servicing / repair
10-38 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
10.8.13 Suction tube
10.8.13.1 Suction rod, rivet, sealing plunger
First remove the head part (1), then the
spreading body (2) of the rivet.
Pull off the grip (1) from the suction rod.
Pull off the tube with the tube clamp (3).
Pull out the sealing plunger (2) in forward
direction.
When installing, ensure proper seating of the
sealing lip; if necessary, correct with
tweezers.