Chapter 7: Error messages
7-6 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
7.1.6 Air detector test
F03 Venous With detuning in positive direction, the achieved change in the venous display is
less than 100 mmHg (7 seconds).
The test detuning is defective (PV_DET, X632/C18) X351/2
Acknowledgement (P_VEN, X351/4) X633L/B5, the change in voltage is
too low
P-venous has not been calibrated
F04 Venous The deviation in the measured value between CPU1 and CPU2 is higher than
±12 mmHg (if P
ven
> 100 mmHg).
Acknowledgement (P_VEN, X351/4) X633L/B5 and X632/C17 measure
different voltage values
P-venous has not been calibrated
F95 Venous System error
Error message Description
F01 Air Detector CPU1 interprets the air detector signal in a different way than does CPU2.
Acknowledgements (LDA1, X351/14) X632/C13 and X633L/C10
recognize different signal levels
F02 Air Detector The air detector alarm is not recognized by CPU2.
Acknowledgement (LDA1, X351/14) X632/C13 not 0 V
Transmission weakening (LDSA, X632/C16) X351/10 not 12 V
F03 Air Detector Air detector clamps acknowledgement (CPU2) activated
(venous line clamp closed).
Acknowledgement (LDA2, X351/6) X632/C14 not 24 V
Clamp control (CLP_CTL, X634/C14) X351/8 not 12 V.
Clamp control (CLP_CTL, X632/C10) X351/8 not 12 V
F04 Air Detector Air detector clamps acknowledgement (CPU1) activated
(venous line clamp closed).
Acknowledgement (LDA2, X351/6) X633L/C13 not 24 V
Clamp control (CLP_CTL, X634/C14) X351/8 not 12 V.
Clamp control (CLP_CTL, X632/C10) X351/8 not 12 V
F05 Air Detector The blood alarm signal has not been cleared (indicates an alarm).
Acknowledgement (BL_AL, X634L/C15) X632/C21 not 12 V
F06 Air Detector Closing of the air detector clamp via the CPU2 control line was not possible.
Clamp control (CLP_CTL, X632/C10) X351/8 not 0 V
Acknowledgement (LDA2, X351/6) X632/C14 not 0 V
F07 Air Detector Opening of the air detector clamp via the CPU2 control line was not possible.
Clamp control (CLP_CTL, X632/C10) X351/8 not 12 V
Acknowledgement (LDA2, X351/6) X632/C14 not 24 V
F08 Air Detector Closing of the air detector clamp via the CPU1 control line was not possible, or
CPU2 acknowledgement is incorrect.
Clamp control (CLP_CTL, X634/C14) X351/8 not 0 V.
Acknowledgement (LDA2, X351/6) X632/C14 not 0 V
Error message Description

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals