Chapter 5: Setup / service program
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 5-15
5.1.3.5 Part 4: Setting the conductivity limit
5.1.3.6 Part 5: Cleaning program info sound
5.1.3.7 Part 6: Setting dialysis parameters
SET CONDUCT. LIMIT
Cd Limit: 12.8 mS/cm
DATA STORED
Conf
key
Audio paused
key
Set the desired value
(12.8 to 14.0 mS/cm)
by pressing the +/– keys.
Esc
key
After
approx. 3 s
+/–
+/–
INFO SOUND ( C-PGM )
Info-Sound: ON
Conf
key
Esc
key
DATA STORED
Info-Sound: OFF
Esc
key
DATA STORED
After
approx. 3 s
After
approx. 3 s
Audio
paused
key
Audio
paused
key
SET DIAL PARAMETERS
AdaptedFlow ON/OFF
Conf
key
see Part 6.1
back to menu ?
SET Flow Parameter see Part 6.2
SET Temp. Parameter
see Part 6.3
SET Na/Bic Parameter
see Part 6.4
Conf
key
Conf
key
Conf
key
Conf
key
Conf
key