Chapter 10: Servicing / repair
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 10-9
10.3.7 Assignment of the hydraulic processing unit (HPU)
Valves
V92 Ventilation valve
V24 Dialyzer inlet valve V98 Rinse valve
V24b Dialyzer outlet valve V99 Rinse valve
V26 Bypass valve V100 Rinse valve
V30 Outlet valve V100* Alternative V100 location if the CDS option is not
provided
V31 Balancing chamber valve 1 V102 Central concentrate delivery valve (optional)
V32 Balancing chamber valve 2 V112 Ventilation valve
V33 Balancing chamber valve 3 (hidden by V31) V117 Concentrate check valve
V34 Balancing chamber valve 4 (hidden by V32) V130 bibag
®
emptying valve (hidden by V188)
V35 Balancing chamber valve 5 V183 Test valve
V36 Balancing chamber valve 6 V188 Evacuation valve
V37 Balancing chamber valve 7 (hidden by V35) V189 Retentate valve
V38 Balancing chamber valve 8 (hidden by V36)
V39 Negative pressure valve
Miscellaneous
V41 Water inlet valve 001 Hydraulic processing unit
V43 Fill valve 002 Overflow tube
V84 Disinfection valve 003 DIASAFE
®
plus connector
V86 Recirculation valve 004 Port to bibag
®
connector
V87 Drain valve 005 Port from bibag
®
connector
V91 Rinse valve V92 Ventilation valve
X B2
Pumpen / pumps Ventile / valves
X C1
CO1 BI
1
X A5
X A1 X A2
X A6 X A4 X A3
N.C.
X A7
Sensoren / sensors
X R1
X 15
X 14
X 13
X 12
X 11
X 10
OCM_NTC4
OCM_LF1 10
NTC109
LF-Bibag 132
Bibag NTC133
Bibag (LP1628)
ZKV PSW
X 2
X 3
X 5
X 6
X 7
X 8
X 9
NTC2
NTC3
FL-Switch
Niveau 6
LF 7
Reedk. 10
P-Dial 1 (9)
X S1
X S2
X S3
X S4
X S5
X S6
X S7
P-Dial 2
N.C.
N.C.
N.C.
Niveau Konz.
NTC308
LF309
X 16
X B1
X 17
X R3
X R2
Bibag SW
V89 Sens.
X 1
CAN
Conc. P. Bic. P.
Kabelbaum / cable harness
BLD
LP 941 WTR / HPU
N.C.
RJ-45