Chapter 9: Calibration / adjustment
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 9-37
9.1.42 Part 15: Blood leak voltage



CALIBRATE BLD
Adjust BLOOD-LEAK
volt.dimn. = 5.0 V
volt. bll. = 5.0 V
DATA STORED
Conf
key
back to menu ?
Adjust DIMNESS
Conf
key
Conf
key
Conf
key
Audio paused
key
DATA STORED
Audio paused
key
Set 5.0 V on the display
by pressing the +/– keys.
Tolerance for dimness voltage: 5 V ±0.3 V.
Note: If values deviate check the glass burette for contamination.
Close the housing; temperature 37 °C; avoid incident light from an external source.
Set 5.0 V on the display
by pressing the +/– keys.
Esc
key
After
approx. 3 seconds
After
approx. 3 seconds
Tolerance for blood-leak voltage: 5 V ±0.2 V.