Chapter 10: Servicing / repair
10-12 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
10.5.3 Brakes
Dismount the brake roller. Remove the guide
roller (1) underneath.
Ensure that only one brake roller model is
installed (flat brake roller / shaped brake
roller).
Detach all brake cables from the screws.
Remove the lock nuts and screws from the
spring plates (1). Short forward cables with
M5x45 flat head screws; long rearward
cables with M5x55 flat head screws.
Remove 4 clamping strips (2) with 2 screws
each.
Remove the brake shaft with brake pedal (1).
The spring plates can each be removed with
1 screw.