Chapter 5: Setup / service program
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 5-71
5.3 Software update
Preparation Switch the device off, dismount the monitor rear, pull out the monitor.
Perform software update Concerning CPU 2, the update is still performed by replacing the
EPROM.
With software version 11.00 or higher, the update of CPU 1 and
MDC-II (monitor board) is performed using the SD-card:
1. P.C.B. LP 1631 (CPU 1): field 3, set DIP switch 7 to ON.
2. P.C.B. LP 1631: insert SD card
3. Turn the device on
4. The update is performed automatically.
P.C.B. LP 1631: While CPU 1 is updated, LED D1 is flashing. As
soon as the update of CPU 1 is complete, D1 will be illuminated.
Then D3 flashes until the update of the MDC-II board is complete.
As soon as the update is complete, D3 will be illuminated.
The flashing frequency of the LEDs D1 and D3 is approx. 2 Hz in
case of an error approx. 6 Hz). In the event of an error, the LEDs D4
to D8 serve for evaluating the error
(see Error code table page 5-72).
5. Turn the device off
6. Re-set DIP switch 7 to OFF
7. The SD-card may be removed or remain inserted