EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>4008 S

Fresenius Medical Care 4008 S Service Manual

Fresenius Medical Care 4008 S
408 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Chapter 4: Specifications
4-4 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
4.5 Electrical supply
Line voltage 100 V AC, 50 – 60 Hz, 16 A
(Medical care cannot be provided in domestic environments with
line/mains voltage of 100 V).
110 V AC, 50 – 60 Hz, 15 A
120 V AC, 50 – 60 Hz, 14 A
127 V AC, 50 – 60 Hz, 14 A
220 V AC, 50 – 60 Hz, 9 A
230 V AC, 50 – 60 Hz, 9 A
240 V AC, 50 – 60 Hz, 9 A
(The decisive criterion is the line voltage, the operating current and the
frequency specified on the type label of the device.)
Power supply
(internal)
+5 V, +0.3 V, short-circuit-proof
+12 V, +0.4 V, short-circuit-proof
+24 V, 0.7 V, short-circuit-proof
Battery Lead-acid battery (maintenance-free)
18 V (= 1 x 12 V + 1 x 6 V) / 3.4 Ah
Supply voltage for audible alarm if power fails for at least one minute.
Battery LP 632 Snap-hat battery (2.8 V / 48 mAh)
Battery LP 1631 Battery CR 1225, Lithium (3 V / 42 mAh)
Power switch Main power switch, all-pole, simultaneous disconnection
4.6 Fuses
Battery 1 x T 3.15 A; power supply unit, fuse in housing foot (rear) / SI5
Power board T 3.15 A Safety fuse
Axial FF 10 A Safety fuse
Heater board T 2.5 A Safety fuse
T 6.3 A Safety fuse
T 16 A Safety fuse

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals