Chapter 10: Servicing / repair
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 10-35
10.8.10 Blood leak detector
Pull out the motor (1) including cable.
After replacement or repair of the flow pump
motor, set the flow rates, check the relief
pressure, and adjust the current increase.
After replacement or repair of the degassing
pump motor, set the degassing and loading
pressures.
Slide the cap (1) on the tubing upwards.
Press in the detent mechanisms (2) and open
the blood leak detector.
Be absolutely sure to avoid fingerprints and
other contaminations on the glass cuvette
during repair.
Remove the glass cuvette (2).
Unscrew the blood leak detector with
2 screws (1).