Chapter 6: TSC / maintenance
6-6 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
6.4 Explanations on the test report – TSC / MA
TSC, explanations on test report
Identification Technician's name:
Surname and last name of technician.
Device type including option(s) / software version:
Device name with possible options or extras. Software version, if
available.
Service report no.:
Number of the service call.
Customer / customer no.:
Final customer's number.
Serial no.:
Serial number indicated on the type label.
Inventory no.:
Inventory number assigned to the device.
Operating hours:
Operating hours, if a time meter is installed.
Equipment code:
Equipment code indicated on the device
(e.g., EC xxx, E-Code xxx).
Re 1 Visual inspections
Operating condition: Off
Re 1.1 The battery fuse accessible (see illustration below) (1) from the outside
corresponds to the indicated value.
Alternative to checking the fuse: If the seal is undamaged, the battery
fuse does not have to be checked.