EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>4008 S

Fresenius Medical Care 4008 S Service Manual

Fresenius Medical Care 4008 S
408 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #275 background imageLoading...
Page #275 background image
Chapter 9: Calibration / adjustment
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 9-55
9.3.1.4 Alternative: Adjustment without the set for air detector calibration
As an alternative, an adjustment is also possible using the set for air
detector calibration (see chapter 9.3.1.1 page 9-53).
Measuring equipment UMED, HMED or pressure gauge,
venous bubble catcher with syringe filled with degassed water or NaCl
solution (see venous bubble catcher in the AD page 9-52).
Measurement point Air Detector
Calibrating the air detector The device must be in the Service mode,
the jumper J1 / P.C.B. LP 450-2 must be set to the calibration position.
Insert the venous bubble catcher into the air detector
(ultrasonic detector); for the moment, do not insert the line into the
venous occlusion clamp.
Fill the venous bubble catcher. The fluid level must be set to
approx. 10 mm above the top edge of the sensor holder.
Turn potentiometer 1 and potentiometer 2 on P.C.B. LP 450-2
clockwise, until the LED D5 and LED D10 on P.C.B. LP 450-2 are
dark.
Slowly (attention: time constant) turn potentiometer 1
counterclockwise, until LED D5 lights.
Slowly (attention: time constant) turn potentiometer 2
counterclockwise, until LED D10 lights (see figure LP 450-2).
After completion of the calibration procedure, set the jumper J1 /
P.C.B. LP 450-2 back to the operation position.
Checking the air detector Lower the fluid level in the venous bubble catcher:
An alarm must be emitted. LED D5 and LED D10 light.
If one or both LEDs are not illuminated, the calibration procedure
must be repeated.
Raise the fluid level in the venous bubble catcher:
It must be possible to clear the alarm; both LEDs must be off.
Remove the venous bubble catcher from the air detector.
In addition, the venous occlusion clamp has to be checked and the
optical detector has to be calibrated and checked
(see chapter 9.3.1.2 page 9-54).

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals