Chapter 9: Calibration / adjustment
9-52 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
Fig.: venous bubble catcher in the AD
Warning
For a calibration of the air detector, the device must be in the Service
mode. The ambient temperature should be between 15 °C and 35 °C.
Warning
The calibration using the set for the air detector calibration is only valid
for devices running with Fresenius tubing systems.
Observe the “use by” date!
Fluid level
syringe filled with degassed water or NaCl
solution