Chapter 3: Installation
3-6 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
3.4 Explanations for the initial start-up report
Identification Technician's name:
Surname and last name of technician.
Device type including option(s) / software version:
Device name with possible options or extras. Software version, if
available.
Service report no.:
Number of the service call.
Customer / customer no.:
Final customer's number.
Serial no.:
Serial number indicated on the type label.
Inventory no.:
Inventory number assigned to the device.
Operating hours:
Operating hours, if a time meter is installed.
Equipment code:
Equipment code indicated on the device
(e.g., EC xxx, E-Code xxx).
Re 1 Preparation
Operating condition: Off
Re 1.1 Check and insert the battery fuse.
Install the battery fuse accessible (1) (see illustration below) after
checking the value.
Attach the seal, if required. Then the battery fuse does not have to be
checked when servicing the device the next time.
Re 1.2 The following closing plugs have to be removed:
– on the overflow tube of the heater block
– on the disinfectant connector
– on the water inlet plate