Chapter 11: Functional description
11-16 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
11.1.1.15 Positive pressure holding test
Test description Check valves V24, V24b and V26 for proper functioning (mechanically).
Test the TMP pressure unit by electronically detuning it in positive
direction.
With the dialysate flow off, apply positive pressure to the balancing
system. The actual value of the dialysate pressure transducer is now
monitored within a defined time period.
Test the pump segment of P97.