Chapter 10: Servicing / repair
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 10-29
– Insert the nut (1) with protective conductor
(2) as shown in the illustration.
– Insert the heating rod.
– Turn the heating rod clockwise hand tight.
– Install the heating rod connector.
– Seals (1) are inserted correctly.
– Replace the heating block cover.
– Tighten the 5 screws with the appropriate
torque
(see chapter 10.4.2 page 10-10).