Chapter 7: Error messages
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 7-33
HPU Error F10 Malfunction of valve MV39
– MV39 defective or activated
– Error on P.C.B. LP 941
HPU Error F11 Malfunction of the test valve (183)
– MV43 defective or activated
– MV183 defective or activated
– Error on P.C.B. LP 941
HPU Error F12 Malfunction of the evacuation valve (188)
– MV188 defective or activated
– Error on P.C.B. LP 941
HPU Error F13 Malfunction of the retentate valve (189)
– MV189 defective or activated
– Error on P.C.B. LP 941
HPU Error F14 Defective component on P.C.B. LP 941
– Error on P.C.B. LP 941
HPU Error F15 Error in the HPU software. Valves are activated incorrectly.
– System error
HPU Error F98 Proceeding to the T1 test is not allowed after restart.
– System error
HPU Error F99 HPU fails without logging off.
– Damaged cable or similar problem
– HPU logged off by CPU1
– CRC error in the transfer HPU CPU1
– The electrical test (VDE) was performed directly after turning the device on.
Turn the device on at least 2 minutes before the test.
Error message Description