Chapter 7: Error messages
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 7-33
HPU Error F10 Malfunction of valve MV39
MV39 defective or activated
Error on P.C.B. LP 941
HPU Error F11 Malfunction of the test valve (183)
MV43 defective or activated
MV183 defective or activated
Error on P.C.B. LP 941
HPU Error F12 Malfunction of the evacuation valve (188)
MV188 defective or activated
Error on P.C.B. LP 941
HPU Error F13 Malfunction of the retentate valve (189)
MV189 defective or activated
Error on P.C.B. LP 941
HPU Error F14 Defective component on P.C.B. LP 941
Error on P.C.B. LP 941
HPU Error F15 Error in the HPU software. Valves are activated incorrectly.
System error
HPU Error F98 Proceeding to the T1 test is not allowed after restart.
System error
HPU Error F99 HPU fails without logging off.
Damaged cable or similar problem
HPU logged off by CPU1
CRC error in the transfer HPU CPU1
The electrical test (VDE) was performed directly after turning the device on.
Turn the device on at least 2 minutes before the test.
Error message Description

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals