Chapter 5: Setup / service program
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 5-31
VENOUS LIMITS ASYM. LIMITS asym. limits: ON asym. limits: ON
asym. limits: OFF
VEN. WINDOW POSITION ven. window position: OFF ven. window position: OFF
ven. window position: ON
SET CENTRAL-DELIVERY NO central-delivery NO central-delivery
central Acid
centr acetate-supply
AutoOFF after AutoON Auto OFF: OFF Auto OFF: ON
Auto OFF: OFF
Init. CAMUS baudrate baudrate: 2400 baud baudrate: 2400 baud
baudrate: 9600 baud
SET BPM Preselected Pressure 160 mmHg 100–290 mmHg 10 mmHg
Upper SYS limit 165 mmHg 95-280 mmHg 5 mmHg
Lower SYS limit 90 mmHg 30-245 mmHg 5 mmHg
Upper DIA limit 100 mmHg 65-240 mmHg 5 mmHg
Lower DIA limit 50 mmHg 10-215 mmHg 5 mmHg
Upper MAP limit 120 mmHg 75-255 mmHg 5 mmHg
Lower MAP limit 70 mmHg 20-230 mmHg 5 mmHg
Upper Pulse limit 120 1/min 45–245 1/min 5 1/min
Lower Pulse limit 60 1/min 20-175 1/min 5 1/min
STORE DEFAULT VALUES Press Audio paused key
Menu item Submenu Default value Value range Resolution
Selectable options