Stress Echo
7-28 Vivid S70 / S60 – User Manual
BC092760-1EN
01
Adding/deleting levels and projections
1. Enter the number of levels and projections in the Grid size
field (see Figure 7-12).
The new grid size is displayed in the Protocol template
preview field.
2. Press New Template to create a new template.
Or
Press Save Template to update the base template.
NOTE: Factory templates cannot be changed.
Scan mode selection
1. From the Scan mode drop-down menu, select a scan mode
to be associated to the actual column (projection).
Display timer(s)
1. Check the box(es) to display timer(s) as specified (see
Figure 7-12).
Start analysis automatically
1. Check Auto start analysis to display the Stress Echo
Analysis screen when the last acquisition is performed.
Crop images
1. Check Crop images to enable display cropping of images
recorded as part of the protocol.
Cropping can be turned off for individual cells of the protocol
acquisition by using the Crop button on Touch panel.
Smart stress
Check Smart stress to store a subset of the image acquisition
settings (e.g. geometry, zoom, gain, compress, reject,
power...etc) for each view in the protocol. Smart Stress enables
to set image acquisition settings for each view at baseline level
and automatically get the same image settings in the
corresponding views in the next levels. If you want to use Smart
stress in Continuous capture, you need to use the arrow buttons
to indicate to the system when you are switching views.