Connect and disconnect probes
Vivid S70 / S60 – User Manual 3-31
BC092760-1EN 01
To disconnect a probe (Small connector - type RS):
1. Freeze the image by pressing Freeze.
2. Press the connector locking lever towards the left to unlock
the connector.
3. Pull the probe and connector straight out of the probe port.
4. Carefully slide the probe and connector away from the
probe port and around the right side of the keyboard.
5. Ensure the cable is free.
6. Be sure that the probe head is clean before placing the
probe in its storage case.
To disconnect a probe (Large connector - type DLP):
1. Rotate the lock handle counter-clockwise to the horizontal
position to unlock the connector.
2. Pull the connector straight out of the probe port.
Transporting probes
When transporting a probe a long distance, store it in its carrying
case.
Storing probes
It is recommended that all probes be stored in the carrying case
provided.
• First place the probe connector into the carrying case.
• Carefully wind the cable into the carrying case.
• Carefully place the probe head into the carrying case. DO
NOT use excessive force or impact the probe face.
Intra Cardiac Ultrasound Catheters
See ‘Intracardiac Ultrasound Catheters’ on page 13-11 for
information.

Table of Contents

Other manuals for GE Vivid S60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60 and is the answer not in the manual?

GE Vivid S60 Specifications

General IconGeneral
BrandGE
ModelVivid S60
CategoryMedical Equipment
LanguageEnglish

Related product manuals