i-4 Vivid S70 / S60 – User Manual
BC092760-1EN
01
Type of protection against electric shock:
• Class I Equipment
Degree of protection against electric shock:
• Type BF Applied part (for Probes marked with BF symbol)
• Type CF Applied part (for ECG, and probes marked with CF
symbol)
• The ECG applied part is protected against defibrillation.
Continuous Operation
System is IPX0.
Footswitch is IPX8.
Transducers are IPX7.
ICE catheters are IPX8.
Class I Equipment
EQUIPMENT in which protection against electric shock does not
rely on BASIC INSULATION only, but also includes an earth
ground. This additional safety precaution prevents exposed
metal parts from becoming LIVE in the event of an insulation
failure.
Type BF Applied part
TYPE BF APPLIED PART providing a specified degree of
protection against electric shock with particular regard to
allowable LEAKAGE CURRENT.
Patient leakage current
Normal condition Single fault condition
d.c. <10 microA <50 microA
a.c. <100 microA <500 microA

Table of Contents

Other manuals for GE Vivid S60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60 and is the answer not in the manual?

GE Vivid S60 Specifications

General IconGeneral
BrandGE
ModelVivid S60
CategoryMedical Equipment
LanguageEnglish

Related product manuals