Probes
13-42 Vivid S70 / S60 – User Manual
BC092760-1EN
01
Coupling gels
In order to ensure optimal transmission of energy between the
patient and probe, a conductive gel or couplant must be applied
liberally to the patient where scanning will be performed.
Refer to the Probe Care Card enclosed in the probe case or to
the Internet link below for the latest list of compatible coupling
gels, cleaners and disinfectans.
Precautions: coupling gels should not contain the following
ingredients as they are known to cause probe damage:
• Methanol, ethanol, isopropanol, or any other alcohol-based
product
• Mineral oil
• Iodine
• Lotions
• Lanoline
• Aloe Vera
• Olive Oil
• Methyl or Ethyl Parabens (para hydroxybenzoic acid)
• Dimethylsilicone
• Polyether glycol based
• Petroleum
CAUTION
Do not apply gel to the eyes. If there is gel contact to the eye,
flush eye thoroughly with water.
CAUTION
Only use GE-recommended gels. Use of unapproved gels may
result in damage to the probe and void the warranty.
http://www3.gehealthcare.com/en/Products/Categories/Ultrasound/Ultrasound_Probes

Table of Contents

Other manuals for GE Vivid S60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60 and is the answer not in the manual?

GE Vivid S60 Specifications

General IconGeneral
BrandGE
ModelVivid S60
CategoryMedical Equipment
LanguageEnglish

Related product manuals