Site requirements
Vivid S70 / S60 – User Manual 3-3
BC092760-1EN 01
Operating Environment
Ensure that there is sufficient air flow around the ultrasound
system when installed or operated.
Environmental requirements
The ultrasound system requires constant maintenance of its
operational environment. Different temperature and humidity
requirements are specified for operation, storage and
transportation.
NOTE: System may be operated at an altitude of up to 3000 meters.
Electromagnetic interferences
The ultrasound system is approved for use in hospitals, clinics,
and other environmentally qualified facilities, in terms of the
prevention of radio wave interference. Operation of the system
in an inappropriate environment can cause electronic
interference to radios and television sets situated near the
medical equipment.
Ensure that the system is protected from electromagnetic
interferences as follows:
• Operate the system at least 4.5 meters (15 feet) away from
equipment that emits strong electromagnetic radiation.
• Shield the system when operating it in the vicinity of radio
broadcasting equipment, if necessary.
Requirement Temperature Humidity Air Pressure
Operational 10–35 °C 30–85% 700–1060 hPa
Storage -20–60 °C 10–95% 700–1060 hPa
Transport -20–60 °C 10–95% 700–1060 hPa

Table of Contents

Other manuals for GE Vivid S60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60 and is the answer not in the manual?

GE Vivid S60 Specifications

General IconGeneral
BrandGE
ModelVivid S60
CategoryMedical Equipment
LanguageEnglish

Related product manuals