Measurements and Analysis
8-50 Vivid S70 / S60 – User Manual
BC092760-1EN
01
4D/Multi-plane LV
4D/multi-plane LV analysis tools
Depending on the data set, the following 4D/multi-plane LV
analysis tools are available on the Vivid S70:
4D Auto LVQ
The 4D Auto LVQ (Automated Left Ventricular Quantification)
tool enables the estimation of the left ventricular volumes and
the ejection fraction in 4D data sets based on automatic border
detection. The tool also enables estimation of left ventricular
mass and strain (only with transthoracic 4D acquisitions). The
automatic border detection is created after placing two points in
an end-diastolic apical view, one at the center of the LV base
and one at the apex.
Data set
Tool 4D Tri-plane Tri-plane TSI
4D Automated LV Quantification page 8-50 - -
Tri-plane LV Volume
measurement
page 8-60 page 8-60 -
Bi-plane LV Volume
measurement
- page 8-63 -
TSI surface model - - page 8-65