4D/Multi-plane LV
Vivid S70 / S60 – User Manual 8-63
BC092760-1EN 01
Rotation of the Volume reconstruction
The volume reconstruction displayed in the Geometric model
can be rotated in any directions.
1. Place the pointer in the Geometric model.
2. Press and hold down Select and use the trackball to rotate
the volume reconstruction.
Bi-plane acquisition
This procedure describes the calculations of the left ventricular
volume from a Bi-plane acquisition. The volume calculation is
based on the Method of disk.
1. In Bi-plane mode, acquire an Apical 4 chamber view in scan
plane 1 (yellow).
2. If required, rotate scan plane 2 to display an Apical
2chamber view.
3. Press Freeze.
4. Using the trackball, scroll through the cineloop to display the
end-diastolic frame.
5. Press Measure.
The Measurement menu is displayed.
6. In the Measurement menu select Volume and Bi-plane.
The Trace tool for the Left ventricular end-diastolic volume
for the Apical 4 chamber view is selected (see Figure 8-32).