Safety
2-24 Vivid S70 / S60 – User Manual
BC092760-1EN
01
Guidance and manufacturer’s decleration - electromagnetic immunity.
Immunity test IEC 60601-1-2 test level Compliance level
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Radiated RF and Proximity fields
from RF wireless
communications equipment
IEC 61000-4-3
3 V/m; 80 MHz to 2.7 GHz
80% AM at 1 kHz
3 V/m; 80 MHz to 2.7 GHz
80% AM at 1 kHz
385 MHz (18 Hz Pulse
Modulation)
27 V/m
450 MHz (FM +/ -5 kHz deviation
1 kHz sine or 18 Hz Pulse
Modulation)
28 V/m
710 MHz (217 Hz PM) 9 V/m
745 MHz (217 Hz PM) 9 V/m
780 MHz (217 Hz PM) 9 V/m
810 MHz (18 Hz PM) 28 V/m
870 MHz (18 Hz PM) 28 V/m
930 MHz (18 Hz PM) 28 V/m
1720 MHz (217 Hz PM) 28 V/m
1845 MHz (217 Hz PM) 28 V/m
1970 MHz (217 Hz PM) 28 V/m
2450 MHz (217 Hz PM) 28 V/m
5240 MHz (217 Hz PM) 9 V/m
5500 MHz (217 Hz PM) 9 V/m
5785 MHz (217 Hz PM) 9 V/m
NOTE: Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.

Table of Contents

Other manuals for GE Vivid S60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60 and is the answer not in the manual?

GE Vivid S60 Specifications

General IconGeneral
BrandGE
ModelVivid S60
CategoryMedical Equipment
LanguageEnglish

Related product manuals