EasyManuals Logo
Home>GE>Medical Equipment>Vivid S60

GE Vivid S60 User Manual

GE Vivid S60
709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #594 background imageLoading...
Page #594 background image
Customize Configuration
12-74 Vivid S70 / S60 – User Manual
BC092760-1EN
01
1. Press View activity report.
The Activity report window is displayed.
2. Adjust the time frame, select the type of studies and users.
3. Select whether to print or save the report as a file.
4. Press Send to create the report.
Figure 12-58. The Activity report window
Unlock patients
If for any reason an examination is not properly finished, the
patient record is locked and cannot be opened again unless it is
unlocked.
To unlock patient records:
1. Press Utility/Config on the Touch panel.
2. Select the Connectivity category and the Other subgroup.
3. In the Other sheet, select the patient record(s) to unlock.
You can search for a specific patient record or a group of
patient record using the searching filters.
4. Select Unlock to unlock the selected patient record(s) or
select Unlock all to unlock all patient records listed.
A Confirmation window is displayed.
5. Select OK.

Table of Contents