September 2007
6-36 General Programming
CALIBRATE RECORDER
Menu
The Calibrate Recorder? menu is used to scale the XMO2’s 4-20 mA
analog output to any range within the maximum calibrated range of
your unit. Refer to the menu map in Figure C-6 on page C-6, and
proceed as follows:
Note: The 4-20 mA analog output range can also be programmed in
the Basic Menu, via the Set Low Input Value (page 5-10) and
Set High Input Value (page 5-11) options.
At the next prompt, enter the oxygen percentage that corresponds to a
4 mA recorder analog output.
At the next prompt, enter the oxygen percentage that corresponds to a
20 mA recorder analog output.
This completes the prompts for the Calibrate Recorder? option, and
the XMO2 returns to the main General menu.
GENERAL MENU
Press [Y] or [Enter] to calibrate
the recorder, or press
[N] to move
to the next General Menu option.
Calibrate Recorder?
mA Output 4 mA Value
To accept the current value, press
[Y] or [Enter]. To change the
current value, use the numeric
keys to enter a new value and
press
[Y] or [Enter] twice.
%O2 [x.xx]:
mA Output 20 mA Value
To accept the current value, press
[Y] or [Enter]. To change the
current value, use the numeric
keys to enter a new value and
press
[Y] or [Enter] twice.
%O2 [xxx.xx]: