Specifications and Factory Data 8-7
September 2007
Default Settings During the original factory calibration, the programming options for
all XMO2 transmitters are set at default values that provide the best
performance in the most common applications. For reference, these
default values are listed in this section.
Basic Menu Defaults The default settings for the default options in the Basic Menu are:
• Quick Offset? - [no default setting]
• View Offset? - [no default setting]
• Set Low Input Value? - [0.00 mA]
• Set High Input Value? - [100.00 mA]
• Resume? - [no default setting]
Any of the above defaults may be changed in the field by the user.
General Menu Defaults The default settings for the options in the General Menu are:
Calibrate System - System Mode
• Response Type - [Damped]
• Pressure Comp - [Off]
• Background Comp - [Off]
• Calibration Mode - [Introduce Gases?]
Calibrate System - Gas Curve
• # Points - [3]
• Point #1 - [0.00 %O2, -400.0 O2 mV]
• Point #2 - [50.00 %O2, 0.0 O2 mV]
• Point #3 - [100.00 %O2, 400.0 O2 mV]
Calibrate System - DriftCal Handler
• Enable DriftCal - [Yes]
• Select number of gases - [One]
• Offset Gas - [20.93 %O2]
• Offset Gas ON for - [0.00 MM:SS]
• Enter Calibration Span - [20.93 %O2]