September 2007
6-40 General Programming
TEMPERATURE INPUT
Option
The Temperature Input option lets you monitor the temperature
sensor input signal. The acceptable range for this signal is 393 mV to
419 mV, with 403 mV to 409 mV being the optimum range.
The current Temperature Input signal is displayed. If the displayed
value is outside the acceptable range, contact the factory for
assistance.
This completes the prompts for the Temperature Input option, and the
XMO2 returns to the Test Inputs? menu.
BRIDGE POWER INPUT
Option
The Bridge Power Input option lets you monitor the Wheatstone
bridge power input signal. The acceptable range for this signal is 6 V
to 14 V.
The current Bridge Power Input signal is displayed.
This completes the prompts for the Bridge Power Input option, and
the XMO2 returns to the Test Inputs? menu.
TEST INPUTS
Press [Y] or [Enter] to view the
Temperature input, or press
[N] to
move to the next Test Inputs?
option.
Temperature Input?
Temperature Input
When ready, press any key to
return to the Test Inputs? menu:
-x.xxx Offset V
TEST INPUTS
Press [Y] or [Enter] to view the
Bridge Power input, or press
[N]
to move to the next Test Inputs?
option.
Bridge Power Input?
Bridge Power Input
When ready, press any key to
return to the Test Inputs? menu:
x.xxx Power V