Advanced Programming 7-21
September 2007
CALIBRATE RECORDER
Menu (cont.)
As the 4 mA and 20 mA calibrations interact with each other, it may
be necessary to repeat the above sequence several times to achieve
and accurate calibration. Accomplish this as follows:
• If an adjustment was required to either of the settings, press [Y] or
[Enter] to repeat the calibration procedure.
• If no adjustment was required to either setting, press [N] to select
another Advanced Menu option or exit the User Program.
SET ERROR HANDLING
Menu
The Set Error Handling? menu is used to individually enable or
disable Terminal and/or mA Output error handling for specific error
conditions. To use the Set Error Handling? menu, refer to the menu
map in Figure C-9 on page C-9 and complete the following steps:
Note: Error handling for the Terminal and/or the mA Output may be
enabled or disabled in total in the Set Error Handling? option
of the General Menu (see Chapter 6, General Programming).
The Set Error Handling? menu includes the following three options:
• Terminal
• mA Output
• Done
Proceed to the appropriate section for programming instructions.
ADVANCED MENU
Continue as indicated below.
Calibrate Recorder?
ADVANCED MENU
Press [N] until this prompt
appears, then press
[Y] or [Enter].
Set Error Handling?