EasyManuals Logo
Home>GE>Transmitter>XMO2

GE XMO2 User Manual

GE XMO2
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #148 background imageLoading...
Page #148 background image
Specifications and Factory Data 8-1
September 2007
Performance
Specifications
Accuracy:
0-1% O
2
range: ±2% of span for
80-100% O
2
and 90-100% O
2
ranges: ±0.2% O
2
all other ranges: ±1% of span
Linearity:
±0.5% of span
Repeatability:
±0.2% of span
Measurement Resolution:
0.01 mA
Stability:
Zero: ±1.0% of span per month (± 2% for 0-1% O
2
range)
Span: ±0.4% of span per month (±0.8% for 0-1% O
2
range)
Response Time:
Fast Response enabled: <5 seconds for 63% of step change
Damped Response enabled: 40 seconds for 63% of step change
Measurement Ranges (typical):
0 to 1% 0 to 25%
0 to 2% 0 to 50%*
0 to 5% 0 to 100%*
0 to 10% 80 to 100%*
0 to 21% 90 to 100%*
* Pressure compensation required
Controlled Sensor Temperature:
Standard: 45°C (113°F)
Optional: 60°C or 70°C (140°F or 158°F)
Atmospheric Pressure Effect:
Standard: ±0.2% of span per mm of Hg
Optional: pressure compensation
Required Sample Gas Flow Rate:
Range: 0.1-2.0 SCFH (50-1,000 cc/min)
Nominal: 1.0 SCFH (500 cc/min)
Sample Gas Flow Rate Effect:
<1% of span, with weatherproof enclosure, background gas
compensation, and 0.1-2.0 SCFH (50-1,000 cc/min) flow rate
Warmup Time:
30 minutes

Table of Contents