5-86 Image Optimization
Map Position
Description
This feature is used to adjust the up/down position of the map.
Operations
Tap [Map Position] on the touch screen to adjust the parameter.
Effects
When the E Average function is enabled or disabled, the elasto curve is displayed
based on different statistical amounts.
Display Format
Description
Adjust the display format of ultrasound image and the Elasto image.
Operation
Tap [H 1:1], [V 1:1], [Full] on the touch screen to adjust.
The system provides 3 types of display format:
H 1:1: right and left display (the real-time ultrasound image appears on the left, and the
elasto image appears on the right);
V 1:1: up down display (the elasto image appears above, and the real-time ultrasound
image appears below).
Full: the elasto image only displayed.
Impacts
Adjust according to the actual situation and obtain a desired analysis through
comparison.
Strain Scale
Description
Adjust the strain scale of the strain bar in different situations.
Operations
Rotate the knob under the [Strain Scale] item on the touch screen to adjust the
parameter.
Effects
The system provides 5 levels of strain scale function: under the same pressure, the
bigger the value, the lower the strain bar.
5.13.1.4 Mass Measurement
Press <Measure> to enter measurement status.
You can measure shell thick, strain ratio, strain-hist, etc.
For details, see [Advanced Volume].
5.13.1.5 Cine Review
Press <Freeze> or open an elastography imaging cine file to enter cine review status.
5.13.2 STE Imaging (Sound Touch Elastography)
Keep the probe still to produce the elastography image in real-time STE mode. The tissue hardness of the
mass can be determined by the image color and brightness. Besides, the relative tissue hardness is
displayed in quantitative manners. STE imaging provides you real elasto modulus for quantification
analysis.
STE imaging is an option.
The SC6-1E probe supports the STE imaging in abdominal exam mode.
The L12-3E, L9-3E, and L14-5WE probes support the STE imaging in small organ (breast, thyroid,
testes) and musculoskeletal exam modes.