13-42 Probes and Biopsy
Hydrogen Peroxide and Peroxyacetic Acid based sterilant:
Chemical
name
Trade name Procedures
22% Hydrogen
Peroxide
4.5%
Peroxyacetic
Acid
MINNCARE LIQUID
DISINFECTANT
Refer to the instructions provided by the solution
manufacturer for details.
Applicable for Canadian region only.
The plastic bracket can be sterilized using MINNCARE LIQUID DISINFECTANT at least 245 times
(11 hours per time) before its safety and performance is affected.
High-pressure steam sterilization (High-pressure steam sterilization is preferred for the metal
guided brackets)
1. Wear a pair of gloves to prevent infection.
2. Clean the needle-guided bracket thoroughly in accordance with the cleaning procedure before
sterilization.
3. Package the needle-guided bracket in accordance with the sterilization requirements of
surgical instruments.
4. Place the packaged needle-guided bracket inside a high-temperature steam sterilizer and
perform sterilization. The sterilization parameters are 121°C and 30 minutes for a gravity
displacement steam sterilizer.
5. Take out the sterilization package after sterilization and dry it in an oven at 60°C for 2030
minutes.
6. Keep the sterilization package together with other sterilized surgical instruments in a sterile
item storage area.
7. Check whether the needle-guided bracket has defects such as deformation and rusting before
use. If such defects exist, the bracket has reached the end of its service life. In this case, stop
using it and contact the Mindray service department.
NOTE:
1. Repeated sterilization with liquid sterilant may degrade the safety and performance
of the plastic needle-guided bracket. Please check the needle-
guided bracket’s
performance periodically.
2. Needle guided brackets are required to verified before each biopsy. If the verification
fails, it indicates that the needle-guided brackets are out of service life. For details
about needle guided bracket verification, please refer to "13.2.4 Biopsy
Menu/Guideline Verification".
13.2.8 Storage and Transportation
1. Do not use the carrying case for storing the needle-guided bracket. If the carrying case is used
for storage, it may become a source of infection.
2. Between examinations, keep the needle-guided bracket in a sterile environment.
3. If the needle-guided bracket is sent to your MINDRAY representative for repair, be sure to
disinfect or sterilize it and keep it in the carrying case to prevent infection.
4. Sterilize the carrying case as necessary.
5. Store or transport the needle-guided bracket under the following ambient conditions:
Ambient temperature: -20°C to 55°C
Relative humidity: 20% to 85% (no condensation)