11-4 DICOM/HL7
Tip:
The AE Title should be the same as the SCU AE Title preset in the server (PACS/RIS/HIS). For
example, if the AE Title of the server preset in the storage server is Storage, and the AE Title of the
accepted SCU is preset as Machine, then in the figure above, the AE Title of Local should be
Machine, and the AE Title of the storage server should be Storage.
11.1.3 Service Preset
The DICOM Service screen is used to set Storage, Print, Worklist, MPPS, Storage Commitment
and Query/Retrieve attributes.
When the system is configured with the DICOM basic function module, and installed with DICOM
Worklist, MPPS, DICOM structured reports and DICOM query/retrieve modules, the corresponding
preset can be found in the DICOM Service screen.
To open the DICOM Service screen:
1. Press the <F10 Setup> key to show the Setup menu.
2. Select [DICOM/HL7] and click [Set DICOM Service] to open the DICOM Service preset screen.
11.1.3.1 Storage Service Preset
1. On the DICOM Service screen, click the [Storage] page tab to enter the Storage page.
2. Select a device and enter the correct AE Title, port, etc.
3. Click [Add] to add the service to the Service List.