EasyManuals Logo
Home>Mindray>Diagnostic Equipment>DC-90

Mindray DC-90 User Manual

Mindray DC-90
391 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #294 background imageLoading...
Page #294 background image
13-14 Probes and Biopsy
Storage and transportation Condition
Store and transport the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
L12-3E
-20~55
30%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
C11-3E -20~55
30%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
SP5-1E -20~55
20%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
DE11-3E
-10~50
30%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
SC6-1E -20~55
20%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
V11-3HE -20~55
20%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
L9-3E -20~55
20%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
SD8-1E -10~50
20%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
L14-5WE -20~55
20%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
P7-3E -20~55
30%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
L20-5E
-20~60
15%~90%RH (no
condensation)
700 hPa ~ 1060 hPa
SC5-1NE
-20~55
20%~95%RH (no
condensation)
700 hPa ~ 1060 hPa
When all examinations for the session have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is in good condition and stored in a
suitable place.
1. To prevent the probe from being damaged, DO NOT store it where it may be exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. When the probe is sent to the MINDRAY Customer Service Department or a sales
representative for repair, be sure to disinfect it and keep it in the carrying case to prevent
infection.
3. Sterilize the carrying case as necessary.