Chapter 8 Specifications and Performance
8.1 Overall Configuration
1-238
8.1.5 TI-LUSU Shutter Unit
Table 8.1-6
Model name TI-LUSU
Input ratings 12 VDC, 1A
8.1.6 AC Adapter for the TI-LUSU Shutter Unit
Table 8.1-7
Model name AD-1260B EA1050E-120
Manufacturer Medi-Power Electronics Inc. (Taiwan) EDAC POWER Electronics Co., Ltd.
Input ratings 100-240 VAC, 1.2 A, 50/60 Hz 100-240 VAC, 1.8 A, 50/60 Hz
Input voltage fluctuation ±10%
Output voltage ratings 12 VDC ±5%
Output current ratings 5 Amax 3.5Amax
Remarks UL listed, GS approved, CE satisfied, and PSE approved product
8.1.7 Power Cord for the AC Adapter of the TI-LUSU Shutter Unit
Be sure to use one of the following power cords:
Table 8.1-8
When used in 100-120 V
region, outside Japan
UL listed detachable power cord set, 3 conductor grounding
(3 conductor grounding Type SVT, No.18 AWG, 3 m long maximum,
rated at 125 VAC minimum)
When used in 220-240 V
region
Detachable power cord set approved according to EU/EN standard,
3 conductor grounding
(3 conductor grounding Type H05VV-F, 3 m long maximum,
rated at 250 VAC minimum)
When used inside Japan
PSE approved detachable power cord set, 3 conductor grounding
(3 conductor grounding Type VCTF 3x0.75mm
2
, 3 m long maximum,
rated at 125 VAC minimum)