RIGOL 6 To Make Measurements
6-10 DS6000 User’s Guide
Press MATH Operate to select "Logic":
Press Formula to select the desired operation formula and the default is AND.
Press Source A and Source B to select the desired channels. The channels
available are CH1, CH2, CH3 and CH4.
Press Invert to turn the invert display of the operation results on or off.
VERTICAL POSITION, VERTICAL SCALE, HORIZONTAL
POSITION and HORIZONTAL SCALE can also be used to adjust the
display of the operation results.
Press Threshold A and use to set the threshold of source A in logic
operation.
Press Threshold B and use to set the threshold of source B in logic
operation.
Press Label to turn on the label modifying interface. You can refer to the
introduction in "
Channel Label" to set the label of the math operation
waveform. Note: The new label can include uppercase English letters (A to Z),
lowercase English letters (a to z), numbers (0 to 9), underlines and spaces and
the length cannot exceed 4 characters.